Human Compiement C5 ELISA Kit

ng/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0061 |
0.0057 |
0.0059 |
|
1.37 |
0.0318 |
0.0300 |
0.0309 |
0.0250 |
4.12 |
0.0813 |
0.0794 |
0.0804 |
0.0745 |
12.35 |
0.2275 |
0.2267 |
0.2271 |
0.2212 |
37.04 |
0.6468 |
0.6343 |
0.6406 |
0.6347 |
111.11 |
1.5860 |
1.5560 |
1.5710 |
1.5651 |
333.33 |
3.2300 |
3.0340 |
3.1320 |
3.1261 |
1000.00 |
4.1306 |
3.9187 |
4.0247 |
4.0188 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(ng/ml) |
17.8 |
92.6 |
287.7 |
16.6 |
85.8 |
261.2 |
|
Standard deviation |
0.7 |
4.5 |
15.2 |
0.7 |
4.8 |
16.9 |
|
Coefficient of variation(%) |
3.7 |
4.9 |
5.3 |
4.5 |
5.6 |
6.5 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 91% to 118% with an overall mean recovery of 101%.
Sample Matrix |
Sample Evaluated |
Range (μg/ml) |
Detectable % |
Mean of Detectable (μg/ml) |
Serum |
30 |
1.79-58.93 |
100 |
25.41 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of Complement C5 in this assay. No medical histories were available for the donors.