Mouse TREM2 Human IL-27 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0253 |
0.0312 |
0.0283 |
|
2.74 |
0.0348 |
0.0328 |
0.0338 |
0.0056 |
8.23 |
0.0466 |
0.0418 |
0.0442 |
0.0160 |
24.69 |
0.0882 |
0.0877 |
0.0880 |
0.0597 |
74.07 |
0.1885 |
0.1928 |
0.1907 |
0.1624 |
222.22 |
0.5228 |
0.4881 |
0.5055 |
0.4772 |
666.67 |
1.3710 |
1.3660 |
1.3685 |
1.3403 |
2000.00 |
3.1040 |
3.1460 |
3.1250 |
3.0968 |
Intra-assay Precision |
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
22 |
22 |
22 |
6 |
6 |
6 |
|
Average(pg/ml) |
35.4 |
192.9 |
626.2 |
39.0 |
202.6 |
679.0 |
Standard Deviation |
2.7 |
13.0 |
47.8 |
1.8 |
7.6 |
34.7 |
Coefficient of Variation(%) |
7.6 |
6.7 |
7.6 |
4.5 |
3.8 |
5.1 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 85% to 107% with an overall mean recovery of 102%.
Sample Matrix |
Sample Evaluated |
Range (ng/ml) |
Detectable % |
Mean of Detectable (ng/ml) |
Serum |
30 |
7959.19-21290.31 | 100 | 14637.59 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of IL-12/IL-23 p40 in this assay. No medical histories were available for the donors.