Human IL1β TR-FRET Detection Kit
Assay Time | 2 hours to overnight at room temperature |
Detection Range | 133 – 20,000 pg/mL |
Limit of Detection (LoD) | 16.8 |
Limit of Quantification (LoQ) | 133.0 |
Specificity | Human IL1β |
Sample Type | Cell culture supernatant, serum, plasma |
Storage | 2-8 °C (lyophilized) / ≤ -60 °C (liquid) |

<table><tr><td>Standard</td><td>Concentration (pg/mL)</td><td>TR-FRET Ratio</td><td>Delta Ratio</td><td>CV%</td></tr><tr><td>STD-7</td><td>20,000.0</td><td>9,307</td><td>9,068</td><td>2.8</td></tr><tr><td>STD-6</td><td>8,000.0</td><td>5,530</td><td>5,292</td><td>2.8</td></tr><tr><td>STD-5</td><td>3,200.0</td><td>2,727</td><td>2,488</td><td>4.6</td></tr><tr><td>STD-4</td><td>1,280.0</td><td>1,399</td><td>1,160</td><td>5.6</td></tr><tr><td>STD-3</td><td>512.0</td><td>720</td><td>482</td><td>3.2</td></tr><tr><td>STD-2</td><td>204.8</td><td>408</td><td>169</td><td>4.0</td></tr><tr><td>STD-1</td><td>81.9</td><td>324</td><td>86</td><td>0.7</td></tr><tr><td>STD-0</td><td>0</td><td>239</td><td>0</td><td>4.6</td></tr></table>
•Low sample consumption (16 µL per well).
•Wide quantifiable range and high reproducibility.
| Parameter | Diluent buffer | RPMI1640 + 10% FBS | DMEM + 10% FBS | 100% FBS |
| Assay range (pg/mL) | 133 – 20,000 pg/mL | |||
| LoD (Std 0 mean + 2×SD) | 16.8 pg/mL | 17.6 pg/mL | 10.6 pg/mL | 51.2 pg/mL |
| LoQ (Std 0 mean + 10×SD) | 133.0 pg/mL | 127.7 pg/mL | 128.5 pg/mL | 189.1 pg/mL |
| Incubation time | 3 hours at room temperature | |||
FAQs
New Products
Instruction