Human IL2 TR-FRET Detection Kit
Assay Time | 2 hours to overnight at room temperature |
Detection Range | 134 – 20,000 pg/mL |
Limit of Detection (LoD) | 28.63 |
Limit of Quantification (LoQ) | 134.0 |
Specificity | Human IL2 |
Sample Type | Cell culture supernatant, serum, plasma |
Storage | ≤ -60 °C / 2-8 °C |

<table><tr><td>Standard</td><td>Concentration (pg/mL)</td><td>TR-FRET Ratio</td><td>Delta Ratio</td><td>CV%</td></tr><tr><td>STD-7</td><td>20,000.0</td><td>28,185</td><td>26,683</td><td>0.7</td></tr><tr><td>STD-6</td><td>7,142.9</td><td>15,971</td><td>14,469</td><td>3.0</td></tr><tr><td>STD-5</td><td>2,551.0</td><td>8,035</td><td>6,533</td><td>2.4</td></tr><tr><td>STD-4</td><td>911.1</td><td>4,175</td><td>2,673</td><td>3.1</td></tr><tr><td>STD-3</td><td>325.4</td><td>2,642</td><td>1,140</td><td>2.4</td></tr><tr><td>STD-2</td><td>116.2</td><td>1,891</td><td>389</td><td>3.1</td></tr><tr><td>STD-1</td><td>41.5</td><td>1,551</td><td>49</td><td>5.8</td></tr><tr><td>STD-0</td><td>0.00</td><td>1,502</td><td>0</td><td>4.9</td></tr></table>
•Low sample consumption (16 µL per well).
•Wide quantifiable range and high reproducibility.
| Parameter | Diluent buffer | RPMI1640 + 10% FBS | DMEM + 10% FBS | 100% FBS |
| Assay range (pg/mL) | 134 – 20,000 pg/mL | |||
| LoD (Std 0 mean + 2×SD) | 28.63 pg/mL | 75.31 pg/mL | 43.59 pg/mL | 197.3 pg/mL |
| LoQ (Std 0 mean + 10×SD) | 134.0 pg/mL | 456.2 pg/mL | 493.9 pg/mL | 1594 pg/mL |
| Incubation time | 2 hours at room temperature | |||