Human beta Amyloid 1-40 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0446 |
0.0451 |
0.0449 |
|
31.25 |
0.0817 |
0.0770 |
0.0794 |
0.0345 |
62.50 |
0.1291 |
0.1349 |
0.1320 |
0.0871 |
125.00 |
0.2772 |
0.2812 |
0.2792 |
0.2343 |
250.00 |
0.5233 |
0.5129 |
0.5181 |
0.4732 |
500.00 |
1.0246 |
1.0388 |
1.0317 |
0.9868 |
1000.00 |
1.8213 |
1.8056 |
1.8135 |
1.7686 |
2000.00 |
3.3130 |
3.3240 |
3.3185 |
3.2736 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
757.9 |
168.5 |
48.0 |
703.8 |
202.7 |
42.7 |
|
Standard deviation |
40.9 |
7.1 |
1.8 |
22.1 |
5.6 |
2.4 |
|
Coefficient of variation(%) |
5.4 |
4.2 |
3.7 |
3.1 |
2.8 |
5.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 84% to 122% with an overall mean recovery of 105%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
Serum |
30 |
n.d.- 907.22 |
37.50 |
198.85 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of beta Amyloid 1-40 in this assay. No medical histories were available for the donors.