Human CCL20/ MIP3 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 8.58 pg/ml (50 μl);12.19 pg/ml (10 μl)
  • Standard Curve Range: 10.97~8000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0028
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
CCL20, also known as MIP-3 alpha, is a chemokine that plays an important role in dendritic cell trafficking and recruitment and activation of T cells. It is upregulated in monocytes, T cells, endothelial cells, epithelial cells, and fibroblasts following inflammatory stimulation. CCL20 signals through CCR6 which is expressed on memory T cells, B cells and bone marrow-derived dendritic cells.
Typical data
pg/ml O.D. Average Corrected
0.00 0.0175 0.0195 0.0185  
10.97 0.0264 0.0238 0.0251 0.0066
32.92 0.0320 0.0311 0.0316 0.0131
98.77 0.0640 0.0629 0.0635 0.0450
296.30 0.2220 0.2240 0.2230 0.2045
888.89 0.9390 0.9138 0.9264 0.9079
2666.67 2.4702 2.4450 2.4576 2.4391
8000.00 3.9750 4.0220 3.9985 3.9800
Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 78.1 395.8 1027.1 82.4 379.2 1127.4
Standard Deviation 2.4 14.8 30.9 4.2 24.1 74.6
Coefficient of Variation(%) 3.1 3.7 3 4 6.4 6.6

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human CCL20 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 121% to 128% with an overall mean recovery of 115%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 n.d.-67.43 20 39.61

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of CCL20 in this assay. No medical histories were available for the donors.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.