Human CCL23/MPIF-1 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0095 |
0.0087 |
0.0091 |
|
4.10 |
0.0230 |
0.0207 |
0.0219 |
0.0128 |
10.24 |
0.0500 |
0.0457 |
0.0479 |
0.0388 |
25.60 |
0.1250 |
0.1321 |
0.1286 |
0.1195 |
64.00 |
0.3903 |
0.3985 |
0.3944 |
0.3853 |
160.00 |
1.3330 |
1.2530 |
1.2930 |
1.2839 |
400.00 |
3.3930 |
3.7000 |
3.5465 |
3.5374 |
1000.00 |
4.3920 |
4.4231 |
4.4076 |
4.3985 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
19.9 |
99.0 |
302.5 |
21.6 |
107.5 |
319.6 |
|
Standard deviation |
1.2 |
3.9 |
9.7 |
1.2 |
4.1 |
13.2 |
|
Coefficient of variation(%) |
5.8 |
3.9 |
3.2 |
5.6 |
3.8 |
4.1 |
- Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
- Inter-assay Precision (Precision between assays)Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human CCL23/MPIF-1 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 93% to 120% with an overall mean recovery of 107%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
Serum |
30 |
131.30-479.56 |
100 |
371.04 |
Serum/Plasma - Thirty samples from apparently healthy volunteers were evaluated for the presence of human CCL23/MPIF-1 in this assay. No medical histories were available for the donors.