Human CCL3/MIP-1 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.05 pg/ml (50 μl);0.52 pg/ml (10 μl)
  • Standard Curve Range: 2.74~2000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0146
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
MIP-1 alpha (macrophage inflammatory protein 1 alpha) is a member of the CC or beta chemokine subfamily that was originally purified from the conditioned media of an LPS-stimulated murine macrophage cell line. MIP-1 alpha acts as a chemoattractant to a variety of cells including monocytes, T cells, B cells and eosinophils. The two human MIP-1 alpha genes arise by duplication/mutation. They code for MIP-1 alpha isoforms CCL3/LD78a and CCL3L1/LD78b, which share 94% amino acid sequence homology. Whereas the human CCL3/LD78a is a single-copy gene, the human CCL3L1/LD78b gene copy number varies within the population. Human CCL3L1/LD78b binds and signals through chemokine receptors CCR1, CCR5. When compared to CCL3/LD78a, CCL3L1/LD78b has higher binding affinity to CCR5, which also functions as a coreceptor for HIV-1 entry. The copy number of CCL3L1 is one of several genetic determinants of HIV-1 susceptibility.
Typical data
pg/ml O.D. Average Corrected
0.00 0.0167 0.0161 0.0164  
2.74 0.0479 0.0428 0.0454 0.0290
8.23 0.0673 0.0832 0.0853 0.0689
24.69 0.2058 0.2010 0.2034 0.1870
74.07 0.6641 0.5979 0.6310 0.6146
222.22 1.4760 1.4170 1.4465 1.4301
666.67 2.5390 2.5340 2.5365 2.5201
2000.00 3.5360 3.5510 3.5435 3.5271
Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
  22 22 22 6 6 6
Average(pg/ml) 55.8 271.9 819.4 33.7 173.5 578.5
Standard Deviation 2.6 17.2 28.5 2.0 13.4 42.3
Coefficient of Variation(%) 4.7 6.3 3.5 6.0 7.7 7.3
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human CCL3/MIP-1 alpha into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 94% to 114% with an overall mean recovery of 102%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 12.28-1051.89 100 224.55
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of CCL3/MIP-1 in this assay. No medical histories were available for the donors.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.