Human CD147/EMMPRIN ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0094 |
0.0094 |
0.0094 |
|
5.49 |
0.0292 |
0.0303 |
0.0298 |
0.0204 |
16.46 |
0.0711 |
0.0747 |
0.0729 |
0.0635 |
49.38 |
0.2033 |
0.2031 |
0.2032 |
0.1938 |
148.15 |
0.5809 |
0.6104 |
0.5957 |
0.5863 |
444.44 |
1.4780 |
1.4830 |
1.4805 |
1.4711 |
1333.33 |
2.8760 |
2.7790 |
2.8275 |
2.8181 |
4000.00 |
3.9430 |
3.8660 |
3.9045 |
3.8951 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
112.4 |
569.7 |
1538.8 |
101.4 |
480.9 |
1376.3 |
|
Standard deviation |
4.0 |
21.3 |
94.7 |
4.6 |
28.0 |
80.6 |
|
Coefficient of variation(%) |
3.6 |
3.7 |
6.2 |
4.5 |
5.8 |
5.9 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 80% to 103% with an overall mean recovery of 86%.The spike recovery was evaluated by spiking 3 levels of human CD147/EMMPRIN into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 80% to 103% with an overall mean recovery of 86%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
Serum |
30 |
139.08-781.71 |
100 |
234.71 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of CD147/EMMPRIN in this assay. No medical histories were available for the donors.