Human CHIT1 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.22 pg/ml (50 μl);0.49 pg/ml (10 μl);
  • Standard Curve Range: 2.74~2000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0186
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
Chitotriosidase encoded by the CHIT1 gene is a member of the chitinase family that is selectively expressed in activated tissue macrophages. It is distinct from another member, known as acidic mammalian chitinase (AMC) encoded by the CHIA gene and expressed mainly in the gastrointestinal tract and lung. Both CHIA and CHIT1 are secreted as 50 kDa proteins. In contrast to CHIA, CHIT1 is not stable under acidic pHs and can be processed into a C-terminally truncated 39 kDa form. CHIT1 is the best biomarker in the monitoring of Gaucher disease among the three most commonly used markers that also include acid phosphatase and angiotensin-converting enzyme (ACE). CHIT1 is also a specific marker of macrophage activation in acute ischemic stroke.
Typical data

pg/ml

O.D.

Average

Corrected

0.00

0.0155

0.0136

0.0146

2.74

0.0232

0.0218

0.0225

0.0080

8.23

0.0381

0.0399

0.0390

0.0245

24.69

0.1039

0.0986

0.1013

0.0867

74.07

0.2901

0.3065

0.2983

0.2838

222.22

0.8336

0.9318

0.8827

0.8682

666.67

2.3258

2.3320

2.3289

2.3144

2000.00

4.0844

4.0230

4.0537

4.0392

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(pg/ml)

47.01

199.00

780.94

46.20

201.75

671.27

Standard Deviation

1.72

9.54

45.28

1.93

5.46

38.00

Coefficient of Variation(%)

3.7

4.8

5.8

4.2

2.7

5.7

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Sample Values
Sample Matrix Sample Evaluated Range (ng/ml) Detectable (%) Mean of Detectable (ng/ml)
Serum 30 0.24-11.50 100 4.04

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of CHIT1 in this assay. No medical histories were available for the donors.

New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.