Human cIAP2 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0139 |
0.0145 |
0.0142 |
|
2.74 |
0.0221 |
0.0234 |
0.0228 |
0.0086 |
8.23 |
0.0328 |
0.0302 |
0.0315 |
0.0173 |
24.69 |
0.0653 |
0.0689 |
0.0671 |
0.0529 |
74.07 |
0.1791 |
0.1718 |
0.1755 |
0.1613 |
222.22 |
0.5172 |
0.5250 |
0.5211 |
0.5069 |
666.67 |
1.4860 |
1.4850 |
1.4855 |
1.4713 |
2000.00 |
3.3930 |
3.4200 |
3.4065 |
3.3923 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
27.3 |
173.6 |
525.6 |
39.3 |
185.7 |
604.2 |
|
Standard deviation |
2.1 |
11.6 |
25.5 |
3.1 |
13.1 |
37.2 |
|
Coefficient of variation(%) |
7.9 |
6.7 |
4.9 |
8.0 |
7.1 |
6.2 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays)Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human cIAP2 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 102% to 119% with an overall mean recovery of 110%.
Sample Matrix |
Sample Evaluated |
Range (Pg/ml) |
Detectable % |
Mean of Detectable (Pg/ml) |
Serum |
30 |
n.d.-4.39 |
33.3% |
1.80 |
Serum/Plasma - Thirty samples from apparently healthy volunteers were evaluated for the presence of human cIAP2 in this assay. No medical histories were available for the donors.