Human CRP ELISA Instructions

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0117 |
0.0092 |
0.0105 |
|
13.72 |
0.0135 |
0.0130 |
0.0133 |
0.0028 |
41.15 |
0.0201 |
0.0218 |
0.0210 |
0.0105 |
123.46 |
0.0482 |
0.0433 |
0.0458 |
0.0353 |
370.37 |
0.1187 |
0.1127 |
0.1157 |
0.1053 |
1111.11 |
0.3375 |
0.3373 |
0.3374 |
0.3270 |
3333.33 |
1.0260 |
0.9954 |
1.0107 |
1.0003 |
10000.00 |
2.7060 |
2.7060 |
2.7060 |
2.6956 |
Intra-assay Precision |
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
22 |
22 |
22 |
6 |
6 |
6 |
|
Average (pg/ml) |
3118.2 |
1092.8 |
240.7 |
3065.9 |
1129.0 |
259.3 |
Standard Deviation |
137.0 |
68.5 |
10.3 |
135.2 |
26.0 |
15.6 |
Coefficient of Variation (%) |
4.4 |
6.3 |
4.3 |
4.4 |
2.3 |
6.0 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human CRP into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 91% to 130% with an overall mean recovery of 109%.
Sample Matrix |
Sample Evaluated |
Range (ng/ml) |
Detectable % |
Mean of Detectable (ng/ml) |
Serum |
30 |
27.88-1049.90 |
100 |
334.69 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of human CRP in this assay. No medical histories were available for the donors.