Human CXCL10/IP-10 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.33 pg/ml (50 μl);1.14 pg/ml (10 μl)
  • Standard Curve Range: 2.74~2000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0026
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
IP-10 was originally identified as an IFN-gamma-inducible gene in monocytes, fibroblasts and endothelial cells. The mouse homolog of human IP-10, CRG-2, shares approximately 67% amino acid sequence identity with human IP-10. The amino acid sequence of IP-10 identified the protein as a member of the CXC chemokine subfamily.
Typical data
pg/ml O.D. Average Corrected
0.00 0.0119 0.0131 0.0125  
2.74 0.0165 0.0173 0.0169 0.0044
8.23 0.0291 0.0303 0.0297 0.0172
24.69 0.0657 0.0682 0.0670 0.0545
74.07 0.1921 0.1924 0.1923 0.1798
222.22 0.6346 0.6597 0.6472 0.6347
666.67 1.8938 1.9571 1.9255 1.9130
2000.00 3.6549 3.8681 3.7615 3.7490
Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 38.3 189.7 714 46.6 195.6 731.2
Standard Deviation 1.6 11.9 33.3 2.7 12.7 41.3
Coefficient of Variation(%) 4.2 6.3 4.7 5.7 6.5 5.7

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human CXCL10 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 81% to 109% with an overall mean recovery of 95%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 4.06-52.70 100 16.04

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of CXCL10 in this assay. No medical histories were available for the donors.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.