Human FGF-19 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0350 |
0.0324 |
0.0337 |
|
15.63 |
0.1448 |
0.1364 |
0.1406 |
0.1069 |
31.25 |
0.2449 |
0.2436 |
0.2443 |
0.2106 |
62.50 |
0.4563 |
0.4771 |
0.4667 |
0.4330 |
125.00 |
0.8452 |
0.8358 |
0.8405 |
0.8068 |
250.00 |
1.6070 |
1.5490 |
1.5780 |
1.5443 |
500.00 |
2.8150 |
2.8420 |
2.8285 |
2.7948 |
1000.00 |
4.1933 |
4.2275 |
4.2104 |
4.1767 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
22.6 |
111.5 |
323.0 |
22.1 |
109.5 |
329.4 |
|
Standard deviation |
1.4 |
4.5 |
11.3 |
0.5 |
2.6 |
12.7 |
|
Coefficient of variation(%) |
6.3 |
4.0 |
3.5 |
2.1 |
2.4 |
3.9 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human FGF-19 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 86% to 107% with an overall mean recovery of 98%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
Serum |
30 |
19.81-294.45 |
100 |
91.01 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of FGF-19 in this assay. No medical histories were available for the donors.