Human FOLR1 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 2.40 pg/ml (50 μl);3.97 pg/ml (10 μl)
  • Standard Curve Range: 46.88~3000 pg/ml
  • Standard Curve Gradient: 7 Points
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0137
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
Folate receptor α (FRα), also known as folate receptor 1 (FOLR1), belongs to the folate receptor family and are both cysteine-rich glycolipid anchor proteins. FRα is encoded by FOLR1 and is a cell surface glycoprotein with a molecular weight of 38-40KD. It was initially discovered as a folic acid-binding protein. It was later found that in addition to transporting folic acid, it is also involved in regulating the proliferation and metastasis of tumor cells. Folic acid is an important nutrient necessary for maintaining cell growth and metabolism, and it plays a key role in the synthesis of DNA and RNA.
typical data

pg/ml

O.D.

Average

Corrected

0.00

0.0158

0.0153

0.0156

 

46.88

0.1271

0.1112

0.1192

0.1036

93.75

0.2292

0.2275

0.2284

0.2128

187.50

0.4572

0.4467

0.4520

0.4364

375.00

0.9516

0.9921

0.9719

0.9563

750.00

1.9200

2.0350

1.9775

1.9620

1500.00

3.9130

3.9100

3.9115

3.8960

3000.00

4.5199

4.5427

4.5313

4.5158

Precision

Intra-assay Precision

 

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(pg/ml)

50.8

260.6

771.5

49.9 

237.4 

720.3 

Standard deviation

3.5

14.6

48.5

2.7 

13.0 

38.8 

Coefficient of variation(%)

6.8

5.6

6.3

5.3

5.5

5.4


Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of human FOLR1 into health human serum sample. The un-spiked serum was used as blank in this experiment.

The recovery ranged from 76% to 120% with an overall mean recovery of 101%.

Sample Values

Sample Matrix

Sample

Evaluated

Range

(pg/ml)

Detectable

%

Mean of Detectable

(pg/ml)

Serum

30

104.68-266.14

100

178.81  


Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of FOLR1 in this assay. No medical histories were available for the donors.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.