Human FOLR1 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0158 |
0.0153 |
0.0156 |
|
46.88 |
0.1271 |
0.1112 |
0.1192 |
0.1036 |
93.75 |
0.2292 |
0.2275 |
0.2284 |
0.2128 |
187.50 |
0.4572 |
0.4467 |
0.4520 |
0.4364 |
375.00 |
0.9516 |
0.9921 |
0.9719 |
0.9563 |
750.00 |
1.9200 |
2.0350 |
1.9775 |
1.9620 |
1500.00 |
3.9130 |
3.9100 |
3.9115 |
3.8960 |
3000.00 |
4.5199 |
4.5427 |
4.5313 |
4.5158 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
50.8 |
260.6 |
771.5 |
49.9 |
237.4 |
720.3 |
|
Standard deviation |
3.5 |
14.6 |
48.5 |
2.7 |
13.0 |
38.8 |
|
Coefficient of variation(%) |
6.8 |
5.6 |
6.3 |
5.3 |
5.5 |
5.4 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human FOLR1 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 76% to 120% with an overall mean recovery of 101%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
Serum |
30 |
104.68-266.14 |
100 |
178.81 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of FOLR1 in this assay. No medical histories were available for the donors.