Human Granzyme A/GZMA ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0136 |
0.0141 |
0.0139 |
|
2.74 |
0.0255 |
0.0239 |
0.0247 |
0.0109 |
8.23 |
0.0392 |
0.0410 |
0.0401 |
0.0263 |
24.69 |
0.0845 |
0.0824 |
0.0835 |
0.0696 |
74.07 |
0.2141 |
0.2112 |
0.2127 |
0.1988 |
222.22 |
0.6253 |
0.5911 |
0.6082 |
0.5944 |
666.67 |
1.7910 |
1.7440 |
1.7675 |
1.7537 |
2000.00 |
3.7830 |
3.7900 |
3.7865 |
3.7727 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
49.6 |
198.7 |
609.8 |
47.5 |
189.0 |
608.9 |
|
Standard deviation |
2.3 |
8.7 |
26.3 |
1.5 |
7.3 |
20.0 |
|
Coefficient of variation(%) |
4.6 |
4.4 |
4.3 |
3.2 |
3.9 |
3.3 |
- Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
- Inter-assay Precision (Precision between assays)Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human Granzyme A into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 81% to 97% with an overall mean recovery of 91%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
Serum |
30 |
62.47-179.55 |
100.0% |
113.58 |
Serum/Plasma - Thirty samples from apparently healthy volunteers were evaluated for the presence of human Granzyme A in this assay. No medical histories were available for the donors.