Human HVEM/TNFRSF14 ELISA Instructions

Key features and details

  • Specification: 96 Test
  • Sensitivity: 1.68 pg/ml (50 μl);1.86 pg/ml (10 μl)
  • Standard Curve Range: 13.72~10000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0107
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
Herpesvirus entry mediator (HVEM), also referred to as TR2 (TNF receptor-like molecule) and ATAR (another TRAF-associated receptor), is a type I membrane protein belonging to the TNF/NGF receptor superfamily. In the TNF superfamily nomenclature, HVEM is referred to as TNFRSF14. Human HVEM cDNA encodes a 283 amino acid (aa) protein with a probable 36 aa signal peptide, a 166 aa extracellular domain, a 23 aa transmembrane region and a 58 aa cytoplasmic region. The extracellular domain of HVEM contains several cysteine-rich repeats characteristic of TNF receptor superfamily members. The short cytoplasmic region lacks a death domain present in some TNF receptor family members, but contains a TRAF (TNF receptor-associated factor) interaction domain. HVEM expression has been detected in peripheral blood T cells, B cells, monocytes and in various tissues enriched in lymphoid cells. The extracellular domain of HVEM has been shown to interact directly with the herpes simplex virus envelope glycoprotein D. Two TNF superfamily ligands, including the secreted TNF‑ beta (lymphotoxin alpha ) and the membrane protein LIGHT (lymphotoxins, exhibits inducible expression, and competes with HSV glycoprotein D for HVEM, a receptor expressed by T lymphocytes), have been shown to be the cellular ligands for HVEM. Besides HVEM, LIGHT can also interact with LT beta R, the receptor for lymphotoxin alpha beta heterotrimer.
Typical Data

pg/ml

O.D.

Average

Corrected

0.00

0.0117

0.0092

0.0105

 

13.72

0.0135

0.0130

0.0133

0.0028

41.15

0.0201

0.0218

0.0210

0.0105

123.46

0.0482

0.0433

0.0458

0.0353

370.37

0.1187

0.1127

0.1157

0.1053

1111.11

0.3375

0.3373

0.3374

0.3270

3333.33

1.0260

0.9954

1.0107

1.0003

10000.00

2.7060

2.7060

2.7060

2.6956

Precision

Intra-assay Precision

Inter-assay Precision

Sample

Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average

(pg/ml)

3118.2

1092.8

240.7

3065.9

1129.0

259.3

Standard

 Deviation

137.0

68.5

10.3

135.2

26.0

15.6

Coefficient 

of 

Variation

(%)

4.4

6.3

4.3

4.4

2.3

6.0


Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays)  Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of human CRP into health human serum sample. The un-spiked serum was used as blank in this experiment.

The recovery ranged from 91% to 130% with an overall mean recovery of 109%. 

Sample Values

Sample

Matrix

Sample 

Evaluated

Range

(ng/ml)

Detectable

%

Mean of Detectable

(ng/ml)

Serum

30

27.88-1049.90

100

334.69


Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of human CRP in this assay. No medical histories were available for the donors.

New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.
© AREX 2024. All Rights Reserved. | 浙ICP备2024113709号-3 Sitemap