Human IFN-γ ELISA Kit

pg/ml | O.D. | Average | Corrected | |
0.00 | 0.0117 | 0.0121 | 0.0119 | |
1.37 | 0.0250 | 0.0282 | 0.0266 | 0.0147 |
4.12 | 0.0462 | 0.0489 | 0.0476 | 0.0357 |
12.35 | 0.1248 | 0.1240 | 0.1244 | 0.1125 |
37.04 | 0.3399 | 0.3393 | 0.3396 | 0.3277 |
111.11 | 0.9201 | 0.8954 | 0.9078 | 0.8959 |
333.33 | 2.0160 | 1.9910 | 2.0035 | 1.9916 |
1000.00 | 3.6940 | 3.6730 | 3.6835 | 3.6716 |
Intra-assay Precision | Inter-assay Precision | |||||
Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
22 | 22 | 22 | 6 | 6 | 6 | |
Average(pg/ml) | 17.2 | 98.2 | 273.3 | 18.9 | 91.8 | 240.1 |
Standard Deviation | 0.5 | 3 | 6.8 | 0.8 | 2.7 | 5.2 |
Coefficient of Variation(%) | 2.8 | 3.2 | 2.5 | 4.2 | 3 | 2.1 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human IFN gamma into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 89% to 113% with an overall mean recovery of 105%
Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
Serum | 30 | n.d.-446.28 | 92 | 93.56 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of IFN-γ in this assay. No medical histories were available for the donors.
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.