Human CKMB ELISA Kit

ng/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0144 |
0.0158 |
0.0151 |
|
0.04 |
0.0332 |
0.0309 |
0.0321 |
0.0170 |
0.12 |
0.0716 |
0.0726 |
0.0721 |
0.0570 |
0.37 |
0.1778 |
0.1780 |
0.1779 |
0.1628 |
1.11 |
0.4794 |
0.4464 |
0.4629 |
0.4478 |
3.33 |
1.3260 |
1.3850 |
1.3555 |
1.3404 |
10.00 |
3.2180 |
3.2230 |
3.2205 |
3.2054 |
30.00 |
4.5063 |
4.4600 |
4.4832 |
4.4681 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(ng/ml) |
0.7 |
3.8 |
13.1 |
0.6 |
3.2 |
11.0 |
|
Standard deviation |
0.1 |
0.2 |
1.0 |
0.1 |
0.1 |
0.3 |
|
Coefficient of variation(%) |
5.6 |
4.6 |
7.3 |
3.6 |
3.9 |
3.1 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human CKMB into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 79% to 119% with an overall mean recovery of 103%.
Sample Matrix |
Sample Evaluated |
Range (ng/ml) |
Detectable % |
Mean of Detectable (ng/ml) |
Serum |
30 |
n.d.-0.06 |
40 |
0.02 |
Serum/Plasma–Thirty samples from apparently healthy volunteers were evaluated for the presence of human CKMB in this assay. No medical histories were available for the donors.