Human IL-3R alpha/CD123 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0491 |
0.0475 |
0.0483 |
|
156.25 |
0.0991 |
0.0987 |
0.0989 |
0.0506 |
312.50 |
0.1488 |
0.1489 |
0.1489 |
0.1006 |
625.00 |
0.2471 |
0.2517 |
0.2494 |
0.2011 |
1250.00 |
0.4688 |
0.4276 |
0.4482 |
0.3999 |
2500.00 |
0.8168 |
0.8312 |
0.8240 |
0.7757 |
5000.00 |
1.5140 |
1.5180 |
1.5160 |
1.4677 |
10000.00 |
2.7410 |
2.7560 |
2.7485 |
2.7002 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
219.2 |
1152.6 |
3499.9 |
194.9 |
1053.1 |
3264.2 |
|
Standard deviation |
12.1 |
67.3 |
170.5 |
12.8 |
58.3 |
145.9 |
|
Coefficient of variation(%) |
5.5 |
5.8 |
4.9 |
6.5 |
5.5 |
4.5 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision
The recovery ranged from 80% to 95% with an overall mean recovery of 86%
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
Serum |
30 |
804.66-2639.86 |
100 |
1600.42 |
Serum/Plasma - Thirty samples from apparently healthy volunteers were evaluated for the presence of human IL-3R alpha/CD123 in this assay. No medical histories were available for the donors.