Human Lipocalin-2/NGAL ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.01 ng/ml (50μl);0.03 ng/ml (10μl);
  • Standard Curve Range: 0.14~100 ng/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 60min
  • Brand:
CAT.NO. : AEHY0166
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
The Lipocalin family comprises a diverse group of mostly secreted soluble proteins that bind hydrophobic ligands and act as transporters, carrying small molecules to specific cells. Lipocalins are related by possessing an 8-stranded beta-barrel structure. Lipocalin-1, also named tear lipocalin (TL), von Ebners gland protein (VEG) and tear pre-albumin, binds a large number of hydrophobic molecules and exhibits cysteine proteinase inhibitor and endonuclear activities. Lipocalin-2, also known as neutrophil gelatinase-associated lipocalin (NGAL), is a component of granules in neutrophils from tissues that are normally exposed to microorganisms and is upregulated during inflammation. Lipocalin-2 can form homodimers and can heterodimerize with the neutrophil gelatinase MMP-9.
Typical data

ng/ml

O.D.

Average

Corrected

0.00

0.0126

0.0115

0.0121

0.14

0.0241

0.0241

0.0241

0.0121

0.41

0.0450

0.0408

0.0429

0.0309

1.23

0.0977

0.0952

0.0965

0.0844

3.70

0.2584

0.2660

0.2622

0.2502

11.11

0.7241

0.7346

0.7294

0.7173

33.33

1.8220

1.8880

1.8550

1.8430

100.00

3.2360

3.2210

3.2285

3.2165

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(ng/ml)

2.3

12.0

41.8

1.7

9.7

38.1

Standard Deviation

0.1

0.3

1.8

0.1

0.5

1.3

Coefficient of Variation(%)

4.7

2.7

4.3

4.5

4.8

3.4

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human Lipocalin-2/NGAL into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 92% to 105% with an overall mean recovery of 97%.
Sample Values
Sample Matrix Sample Evaluated Range (μg/ml) Detectable (%) Mean of Detectable (μg/ml)
Serum 30 11.47-84.79 100 49.12

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of Lipocalin-2/NGAL in this assay. No medical histories were available for the donors.

New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.