Human MERTK ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.03 ng/ml (50 μl);0.41 ng/ml (10 μl)
  • Standard Curve Range: 0.27~200 ng/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 60min
  • Brand:
CAT.NO. : AEHY0189
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
Tyrosine-protein kinase Mer(MERTK) is a member of the TYRO3/AXL/MER (TAM) receptor kinase family and encodes a transmembrane protein with two fibronectin type-III domains, two Ig-like C2-type (immunoglobulin-like) domains, and one tyrosine kinase domain. Following activation by ligand, interacts with GRB2 or PLCG2 and induces phosphorylation of MAPK1, MAPK2, FAK/PTK2 or RAC1. MERTK signaling plays a role in various processes such as macrophage clearance of apoptotic cells, platelet aggregation, cytoskeleton reorganization and engulfment. Functions in the retinal pigment epithelium (RPE) as a regulator of rod outer segments fragments phagocytosis. Plays also an important role in inhibition of Toll-like receptors (TLRs)-mediated innate immune response by activating STAT1, which selectively induces production of suppressors of cytokine signaling SOCS1 and SOCS3.
Typical data

ng/ml

O.D.

Average

Corrected

0.00

0.0114

0.0112

0.0113

0.27

0.0191

0.0184

0.0188

0.0075

0.82

0.0343

0.0338

0.0341

0.0228

2.47

0.0751

0.0732

0.0742

0.0629

7.41

0.1987

0.2044

0.2016

0.1903

22.22

0.4883

0.5378

0.5131

0.5018

66.67

1.4530

1.5690

1.5110

1.4997

200.00

3.4070

3.4100

3.4085

3.3972

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(ng/ml)

3.7

21.9

76.7

3.0

17.4

58.8

Standard Deviation

0.2

0.9

4.1

0.2

1.1

4.5

Coefficient of Variation(%)

5.5

4.0

5.3

5.6

6.1

7.7

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human MERTK into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 70% to 105% with an overall mean recovery of 94%.
Sample Values
Sample Matrix Sample Evaluated Range (ng/ml) Detectable (%) Mean of Detectable (ng/ml)
Serum 30 37.02-124.26 100 56.4

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of MERTK in this assay. No medical histories were available for the donors.

New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.
© AREX 2024. All Rights Reserved. | 浙ICP备2024113709号-3 Sitemap