Human MMP-7 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0753 |
0.0815 |
0.0784 |
|
312.50 |
0.1392 |
0.1352 |
0.1372 |
0.0588 |
625.00 |
0.1928 |
0.1952 |
0.1940 |
0.1156 |
1250.00 |
0.3001 |
0.2910 |
0.2956 |
0.2172 |
2500.00 |
0.4927 |
0.4717 |
0.4822 |
0.4038 |
5000.00 |
0.8668 |
0.8596 |
0.8632 |
0.7848 |
10000.00 |
1.6280 |
1.5980 |
1.6130 |
1.5346 |
20000.00 |
2.9050 |
2.7370 |
2.8210 |
2.7426 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(pg/ml) |
490.3 |
2109.8 |
6708.0 |
456.7 |
2055.5 |
6243.6 |
|
Standard deviation |
30.3 |
81.7 |
274.3 |
25.6 |
94.3 |
290.7 |
|
Coefficient of variation(%) |
6.2 |
3.9 |
4.1 |
5.6 |
4.6 |
4.7 |
- Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
- Inter-assay Precision (Precision between assays)Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human MMP-7 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 82% to 108% with an overall mean recovery of 98%.
Sample Matrix |
Sample Evaluated |
Range (Pg/ml) |
Detectable % |
Mean of Detectable (Pg/ml) |
Serum |
30 |
1696.59-7447.73 |
100 |
4027.16 |
Serum/Plasma - Thirty samples from apparently healthy volunteers were evaluated for the presence of human MMP-7 in this assay. No medical histories were available for the donors.