Human/Mouse BDNF ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0195 |
0.0195 |
0.0195 |
|
1.37 |
0.0241 |
0.0238 |
0.0240 |
0.0045 |
4.12 |
0.0410 |
0.0387 |
0.0399 |
0.0204 |
12.35 |
0.0961 |
0.0886 |
0.0924 |
0.0729 |
37.04 |
0.2670 |
0.2588 |
0.2629 |
0.2434 |
111.11 |
0.7864 |
0.7770 |
0.7817 |
0.7622 |
333.33 |
1.9080 |
1.9160 |
1.9120 |
1.8925 |
1000.00 |
3.3430 |
3.3080 |
3.3255 |
3.3060 |
Intra-assay Precision |
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
22 |
22 |
22 |
6 |
6 |
6 |
|
Average(pg/ml) |
26.4 |
132.7 |
428.3 |
26.0 |
132.0 |
417.2 |
Standard Deviation |
0.9 |
5.5 |
28.5 |
0.8 |
5.2 |
22.7 |
Coefficient of Variation(%) |
3.4 |
4.1 |
6.7 |
3.1 |
4.0 |
5.4 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 90% to 118% with an overall mean recovery of 104%.
Sample Matrix |
Sample Evaluated |
Range (ng/ml) |
Detectable (%) |
Mean of Detectable (ng/ml) |
Serum |
30 |
12.32-21.28 |
100 |
16.8 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of BDNF in this assay. No medical histories were available for the donors.