Human PAI-1/Serpin E1 ELISA Kit
CAT.NO. : AEHY0160
RMB Please choose
RMB Please choose
Size:
Trail, Bulk size or Custom requests Please contact us
Background
The human serpin superfamily consists of at least 35 members that target not only serine proteases, but also selected cysteine proteases and non-protease proteins. Serpins bind the protease active site resulting in a major conformational rearrangement that traps the enzyme in a covalent acyl-enzyme intermediate. As protease inhibitors, serpins have an array of functions including regulating blood clotting, the complement pathway, extracellular matrix remodeling, and cell motility. They are also involved in activities that extend beyond their ability to inhibit proteases. For instance, they may also regulate blood pressure, angiogenesis, or act as storage/transport proteins.
Typical data
|
ng/ml |
O.D. |
Average |
Corrected |
|
|
0.00 |
0.0073 |
0.0071 |
0.0072 |
|
|
0.05 |
0.0128 |
0.0128 |
0.0128 |
0.0056 |
|
0.16 |
0.0240 |
0.0223 |
0.0232 |
0.0160 |
|
0.49 |
0.0578 |
0.0546 |
0.0562 |
0.0490 |
|
1.48 |
0.1659 |
0.1523 |
0.1591 |
0.1519 |
|
4.44 |
0.5147 |
0.4948 |
0.5048 |
0.4976 |
|
13.33 |
1.5870 |
1.5630 |
1.5750 |
1.5678 |
|
40.00 |
3.8260 |
3.8630 |
3.8445 |
3.8373 |
Precision
|
Intra-assay Precision |
Inter-assay Precision |
|||||
|
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
|
|
Average(ng/ml) |
0.9 |
4.1 |
12.3 |
0.8 |
3.8 |
12.2 |
|
Standard Deviation |
0.0 |
0.1 |
0.4 |
0.0 |
0.2 |
0.3 |
|
Coefficient of Variation(%) |
3.1 |
2.8 |
3.0 |
4.7 |
4.2 |
2.7 |
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human PAI-1/Serpin E1 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 85% to 101% with an overall mean recovery of 94%.
The recovery ranged from 85% to 101% with an overall mean recovery of 94%.
Sample Values
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 98.03-475.14 | 100 | 264.19 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of PAI-1/Serpin E1 in this assay. No medical histories were available for the donors.