Human Total HSP60 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0310 |
0.0327 |
0.0319 |
|
4.10 |
0.0436 |
0.0462 |
0.0449 |
0.0131 |
10.24 |
0.0678 |
0.0691 |
0.0685 |
0.0366 |
25.60 |
0.1196 |
0.1236 |
0.1216 |
0.0898 |
64.00 |
0.2535 |
0.2621 |
0.2578 |
0.2260 |
160.00 |
0.5522 |
0.5783 |
0.5653 |
0.5334 |
400.00 |
1.2460 |
1.2700 |
1.2580 |
1.2262 |
1000.00 |
2.5840 |
2.5540 |
2.5690 |
2.5372 |
Intra-assay Precision |
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
22 |
22 |
22 |
6 |
6 |
6 |
|
Average(pg/ml) |
26.6 |
99.0 |
277.0 |
27.8 |
101.2 |
284.9 |
Standard Deviation |
2.3 |
5.6 |
21.7 |
1.8 |
5.3 |
14.5 |
Coefficient of Variation(%) |
8.5 |
5.6 |
7.8 |
6.3 |
5.2 |
5.1 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 66% to 73% with an overall mean recovery of 69%.
Sample Matrix |
Sample Evaluated |
Range (ng/ml) |
Detectable % |
Mean of Detectable (ng/ml) |
Serum |
30 |
n.d.-20.90 | 50 | 8.98 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of IL-12/IL-23 p40 in this assay. No medical histories were available for the donors.