Human tPA ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.03 ng/ml (50 μl);0.10 ng/ml (10 μl);
  • Standard Curve Range: 0.14~100 ng/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 60min
  • Brand:
CAT.NO. : AEHY0158
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
PLAT, also known as tissue-type plasminogen activator (tPA), is a secreted serine protease synthesized by endothelial cells (1). The partially active single chain can be further processed to full activity by plasmin, tissue kallikrein or Factor Xa. Active PLAT converts plasminogen to plasmin, a fibrinolytic protease, by hydrolyzing an Arg-Val peptide bond in plasminogen. Unusually high levels of tPA activity can result in excessive bleeding, and low levels of tPA activity can result in thrombosis or embolism. Human PLAT contains 4 domains; the N-terminal fibronectin type-1 domain, an epidermal growth factor-like domain, two kringle domains and a serine protease catalytic domain.
Typical data

ng/ml

O.D.

Average

Corrected

0.00

0.0058

0.0064

0.0061

0.14

0.0107

0.0096

0.0102

0.0041

0.41

0.0195

0.0181

0.0188

0.0127

1.23

0.0457

0.0450

0.0454

0.0393

3.70

0.1406

0.1261

0.1334

0.1273

11.11

0.4191

0.3776

0.3984

0.3923

33.33

1.1900

1.1700

1.1800

1.1739

100.00

2.8370

2.8780

2.8575

2.8514

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(ng/ml)

1.8

9.0

29.2

1.8

9.4

31.7

Standard Deviation

0.1

0.4

1.5

0.1

0.4

1.8

Coefficient of Variation(%)

5.7

4.6

5.1

5.2

3.9

5.6

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human tPA into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 97% to 115% with an overall mean recovery of 104%.
Sample Values
Sample Matrix Sample Evaluated Range (ng/ml) Detectable (%) Mean of Detectable (ng/ml)
Serum 30 3.24-20.65 100 9.25

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of tPA in this assay. No medical histories were available for the donors.

New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.