Human VE Cadherin ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0176 |
0.0176 |
0.0176 |
|
0.14 |
0.0228 |
0.0224 |
0.0226 |
0.0050 |
0.41 |
0.0323 |
0.0328 |
0.0326 |
0.0150 |
1.23 |
0.0583 |
0.0614 |
0.0599 |
0.0423 |
3.70 |
0.1408 |
0.1456 |
0.1432 |
0.1256 |
11.11 |
0.3962 |
0.3948 |
0.3955 |
0.3779 |
33.33 |
1.0930 |
1.1360 |
1.1145 |
1.0969 |
100.00 |
2.8490 |
2.8830 |
2.8660 |
2.8484 |
Intra-assay Precision |
|
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
||
Average(ng/ml) |
37.7 |
10.4 |
2.0 |
32.5 |
9.0 |
1.8 |
|
Standard deviation |
1.8 |
0.5 |
0.1 |
0.9 |
0.3 |
0.1 |
|
Coefficient of variation(%) |
4.9 |
4.7 |
3.5 |
2.7 |
3.3 |
3.3 |
- Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
- Inter-assay Precision (Precision between assays)Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human VE Cadherin into health rat serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 79% to 122% with an overall mean recovery of 98% .
Sample Matrix |
Sample Evaluated |
Range (ng/ml) |
Detectable % |
Mean of Detectable (ng/ml) |
Serum |
30 |
n.d.-0.18 |
18.8 |
0.13 |
Serum/Plasma - Thirty samples from apparently healthy human were evaluated for the presence of human VE Cadherin in this assay. No medical histories were available for the donors.