Human VEGF/VEGFA ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0275 |
0.0264 |
0.0270 |
|
31.25 |
0.1190 |
0.1082 |
0.1136 |
0.0867 |
62.50 |
0.1991 |
0.2040 |
0.2016 |
0.1746 |
125.00 |
0.3159 |
0.2863 |
0.3011 |
0.2742 |
250.00 |
0.5521 |
0.6045 |
0.5783 |
0.5514 |
500.00 |
1.1560 |
1.2260 |
1.1910 |
1.1641 |
1000.00 |
2.3170 |
2.2130 |
2.2650 |
2.2381 |
2000.00 |
3.6280 |
3.4990 |
3.5635 |
3.5366 |
Intra-assay Precision | Inter-assay Precision | |||||
Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
22 | 22 | 22 | 6 | 6 | 6 | |
Average(pg/ml) | 170.7 | 765.3 | 2207.8 | 168.7 | 808.5 | 2407.2 |
Standard Deviation | 8.6 | 42.2 | 87.7 | 7.6 | 25 | 164.5 |
Coefficient of Variation(%) | 5.0 | 5.5 | 4.0 | 4.5 | 3.1 | 6.8 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human VEGF/VEGFA into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 94% to 119% with an overall mean recovery of 105%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
Serum |
30 |
25.21-450.32 |
100.0 |
185.52 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of VEGF in this assay. No medical histories were available for the donors.