Mouse Coagulation Factor XII ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 2.35 pg/ml (10 μl)
  • Standard Curve Range: 13.72~10000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEM0041
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
Coagulation factor XII (F12) is a serum glycoprotein that plays multiple roles in initiating blood coagulation, fibrinolysis, and the production of bradykinin and angiotensin.
typical data

pg/ml

O.D.

Average

Corrected

0.00

0.0164

0.0142

0.0153

 

13.72

0.0253

0.0224

0.0239

0.0086

41.15

0.0286

0.0276

0.0281

0.0128

123.46

0.0472

0.0489

0.0481

0.0328

370.37

0.1131

0.1289

0.1210

0.1057

1111.11

0.3938

0.4223

0.4081

0.3928

3333.33

1.3840

1.4040

1.3940

1.3787

10000.00

3.3980

3.3740

3.3860

3.3707

Precision

Intra-assay Precision

 

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(pg/ml)

186.2

1021.9

3453.9

143.3

846.7

2856.7

Standard deviation

5.7

43.2

230.7

6.7

56.7

113.3

Coefficient of variation(%)

3.0

4.2

6.7

5.5

6.6

4.0


Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of Mouse Coagulation Factor XII into health Mouse serum sample. The un-spiked serum was used as blank in this experiment.

The recovery ranged from 93% to 106% with an overall mean recovery of 100%.

Sample Values

Sample Matrix

Sample

Evaluated

Range

(ug/ml)

Detectable

%

Mean of Detectable

(ug/ml)

Serum

30

13.00-60.31

100

33.76


Serum/Plasma – Thirty samples from apparently healthy mice were evaluated for the presence of Coagulation Factor XII in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.
© AREX 2024. All Rights Reserved. | 浙ICP备2024113709号-3 Sitemap