Key features and details

  • Specification: 96 Test
  • Sensitivity: 10.2 pg/ml (10 μl)
  • Standard Curve Range: 13.72~10000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEM0018
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
CXCL2/GRO beta, also called MIP-2 in mouse and CINC-3 in rat, is a member of the CXC chemokine family. Human CXCL2/GRO beta is 107 amino acids (aa) in length with a predicted molecular weight of 11 kDa. The mouse and rat orthologs share 70% and 71% aa sequence identity with the human protein, respectively. N-terminal aa 1-4 of CXCL2/GRO beta can be post-translationally cleaved which confers enhanced hematopoietic bioactivity. CXCL2/GRO beta is produced by a variety of cell types including monocytes and macrophages at sites of inflammation and is chemotactic for granulocytes, including neutrophils.
Typical data
pg/ml O.D. Average Corrected
0.00 0.0130 0.0122 0.0126  
13.72 0.0151 0.0142 0.0147 0.0021
41.15 0.0210 0.0210 0.0210 0.0084
123.46 0.0381 0.0411 0.0396 0.0270
370.37 0.1251 0.1312 0.1282 0.1156
1111.11 0.5023 0.5012 0.5018 0.4892
3333.33 1.7271 1.8113 1.7692 1.7566
10000.00 3.9910 4.0931 4.0421 4.0295
Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 160.2 1021.4 3095.3 153.9 1034.9 3264.4
Standard Deviation 8.9 64.2 148.8 9.0 55.2 180.8
Coefficient of Variation(%) 5.6 6.3 4.8 5.9 5.3 5.5

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery
The spike recovery was evaluated by spiking 3 levels of mouse CXCL2/MIP-2 into health mouse serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 89% to 114% with an overall mean recovery of 101%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 n.d. 0 n.d.

Serum/Plasma – Thirty samples from apparently healthy mice were evaluated for the presence of CXCL2/MIP-2 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.