Mouse IL-9 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.10 pg/ml (10 μl)
  • Standard Curve Range: 1.37~1000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEM0022
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
Mouse interleukin 9 (IL-9) was originally identified as T cell-derived T cell growth factor III/P40 that could support the long term growth of certain mouse T helper clones in the absence of antigen or antigen-presenting cells. Human IL-9 was independently cloned as a novel growth factor that is mitogenic for the human megakaryoblastic leukemic cell line, M07e. Human and mouse IL-9 share 56% identity at the amino acid level. Although mouse IL-9 is active on human cells, human IL-9 is not active on mouse cells.
Typical data
pg/ml O.D. Average Corrected
0.00 0.0244 0.0248 0.0246  
1.37 0.0455 0.0463 0.0459 0.0213
4.12 0.0790 0.0843 0.0817 0.0571
12.35 0.1971 0.1957 0.1964 0.1718
37.04 0.4988 0.4858 0.4923 0.4677
111.11 1.1970 1.1910 1.1940 1.1694
333.33 2.1870 2.1790 2.1830 2.1584
1000.00 2.8400 2.8820 2.8610 2.8364
Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 20.5 100.5 247.8 20.0 104.3 314.7
Standard Deviation 1.1 7.1 20.1 0.5 2.5 17.5
Coefficient of Variation(%) 5.2 7 8.1 2.7 2.4 5.6

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery
The spike recovery was evaluated by spiking 3 levels of mouse IL-9 into health mouse serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 105% to 125% with an overall mean recovery of 117%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 n.d.-2.03 93 0.56

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated in this assay. Donor medical histories were not available.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.