Mouse PD-L1 ELISA Kit

pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0578 |
0.0568 |
0.0573 |
|
1.37 |
0.0794 |
0.0801 |
0.0798 |
0.0225 |
4.12 |
0.1089 |
0.0990 |
0.1040 |
0.0467 |
12.35 |
0.1731 |
0.1708 |
0.1720 |
0.1147 |
37.04 |
0.3835 |
0.3865 |
0.3850 |
0.3277 |
111.11 |
0.9157 |
0.9152 |
0.9155 |
0.8582 |
333.33 |
2.1664 |
2.1622 |
2.1643 |
2.1070 |
1000.00 |
3.9609 |
3.9586 |
3.9598 |
3.9025 |
Intra-assay Precision |
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
22 |
22 |
22 |
6 |
6 |
6 |
|
Average(pg/ml) |
24.1 |
109.5 |
371.6 |
24.4 |
127.7 |
359.1 |
Standard Deviation |
0.8 |
6.8 |
8.5 |
1.0 |
3.9 |
6.9 |
Coefficient of Variation(%) |
3.3 |
6.2 |
2.3 |
4.0 |
3.1 |
1.9 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 86% to 116% with an overall mean recovery of 107%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable (%) |
Mean of Detectable (pg/ml) |
Serum |
30 |
122.17-529.36 |
100 |
341.58 |