Mouse TNF-α ELISA Kit

pg/ml | O.D. | Average | Corrected | |
0.00 | 0.0121 | 0.0111 | 0.0116 | |
1.37 | 0.0242 | 0.0223 | 0.0233 | 0.0117 |
4.12 | 0.0482 | 0.0444 | 0.0463 | 0.0347 |
12.35 | 0.1134 | 0.1092 | 0.1113 | 0.0997 |
37.04 | 0.3165 | 0.3032 | 0.3099 | 0.2983 |
111.11 | 0.8042 | 0.7996 | 0.8019 | 0.7903 |
333.33 | 1.9672 | 1.9564 | 1.9618 | 1.9502 |
1000.00 | 4.1683 | 4.2715 | 4.2199 | 4.2083 |
Intra-assay Precision | Inter-assay Precision | |||||
Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
22 | 22 | 22 | 6 | 6 | 6 | |
Average(pg/ml) | 16.1 | 80.4 | 271.7 | 14 | 75.2 | 236.7 |
Standard Deviation | 1.1 | 4.8 | 16 | 0.3 | 4 | 10.8 |
Coefficient of Variation(%) | 6.5 | 5.9 | 5.9 | 2 | 5.3 | 4.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of mouse TNF alpha into mouse serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 107% to 132% with an overall mean recovery of 116%.
Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
Serum | 30 | n.d.-0.28 | 33 | 0.16 |
Serum/Plasma – Thirty samples from apparently healthy mice were evaluated for the presence of TNF alpha in this assay. No medical histories were available for the donors.
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.