Rat CRP/C-Reactive Protein ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 27.24 pg/ml (50 μl);63.91 pg/ml (10 μl)
  • Standard Curve Range: 81.92~20000 pg/ml
  • Standard Curve Gradient: 7 Points
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AER0009
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
*产品价格可能会有所调整,请以品牌方官网实时更新的价格为准,以确保准确性。
Product Details
Background
C-Reactive Protein (CRP), also known as Pentraxin 1, is a secreted pentameric protein that functions as a sensor and activator for the innate immune response. In humans, it is a major acute-phase protein; its circulating concentration is dramatically elevated at the onset of inflammation. In mice, however, serum CRP levels increase only slightly during inflammation, and the analogous acute phase role is filled by Pentraxin 2. CRP binds, opsonizes, and induces the phagocytosis of bacteria and apoptotic cells. It regulates activation of the classical complement pathway by binding several proteins in the complement cascade as well as Fc gamma RI, Fc gamma RIIA, and Fc gamma RIIB on macrophages and dendritic cells. It also promotes dendritic cell maturation and humoral immunity. In cardiovascular disease, CRP binds to oxidized LDL, exacerbates tissue damage in myocardial infarction, and inhibits the repair of injured vascular endothelium.
Typical data

pg/ml O.D. Average Corrected
0.00 0.0248 0.0257 0.0253  
81.92 0.0392 0.0371 0.0382 0.0232
204.80 0.0777 0.0792 0.0785 0.0635
512.00 0.1553 0.1399 0.1476 0.1326
1280.00 0.3510 0.3545 0.3528 0.3378
3200.00 0.9028 0.8789 0.8909 0.8759
8000.00 2.0410 2.1280 2.0845 2.0695
20000.00 3.6460 3.9210 3.7835 3.7685
Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 486.6 2033.5 6246.1 483.8 2222.2 6448.2
Standard Deviation 25.5 121.9 330.3 15.4 112.6 123.97
Coefficient of Variation(%) 5.2 6 4.8 3.2 5.1 1.9

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery
The spike recovery was evaluated by spiking 3 levels of rat CRP into health rat serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 74% to 137% with an overall mean recovery of 103%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 96.31-1708.35 100 398.28

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of IL-8 in this assay. No medical histories were available for the donors.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.